Physical specifications:
Display LED Display
Mode: pneumatically powered and electrically controlled system
Trolley: Fitted with 4 nos anti-static rubber castors; two of which are lockable for braking and easy maneuverability with foot operated brake provisions
Technical Specifications:
Gas requirement: Medical oxygen with a pressure ranging from 280~600kpa
safety valve <12.5 kPa
Respiratory Rate 6~60 times/ min
Oxygen concentration output <50%
minute volume ≥18L
modes of ventilation A/C, IPPV, SIPPV, IMV, SIMV, PEET. MANUAL
PEEP: 0.1 ~ 1.0 kPa
Ptr: -4 ~ 10 hPa
IMV frequency: 1~12 bpm
I/E Ratio: 1:1.5; 1:2.0; 1:2.5; 1:3.0
Tidal Volume 0-2000ml
Inspiratory Plateau: 0~1s
Sigh: one deep breath per 80 controlled respirations, the inspiration time is 1.5 times of the setting point
Power Supply Alarm "Ac/dc power supply are after failing to send out alarm immediately
Alarm Time: keep >120s"
The airway pressure continues to be higher than 15 hPa ±1 hPa for 15s±1s, then the machine will raise an audible alarm,
the pressure will be displayed in red and the continuous high pressure red alarm wording is displayed on the screen of the anesthetic respirator.