IN-C041 Hospital First-Aid Multiparameter patient monitor

Item No.: IN-C041
Siz:12.1" color TFT screen
Display:Multi-channel display
Design:Anti-defibrillation design
Interface:Big font interface
Analysis:ST and arrhythmia analysis
Standard:ECG, Respiration, NIBP, SpO2, Pulse Rate, Temperature
Description

Technical Specifications:

 

Lead Mode 5 Leads (R, L, F, N, C or RA, LA, LL, RL,V) 
Lead selection I, II, III, avR, avL, avF, V
Wavef orm 2 ch
Lead mode 3 Leads ( R, L, F or RA, LA, LL) 
Lead selection I, II, III,
Wave form 1 ch
Gain 2.5mm/mV, 5.0mm/mV, 10mm/mV,20mm/mV, auto
HR and Alarm Range Adult 15 ~ 300 bpm


Neo/Ped 15 ~ 350 bpm
Accuracy ±1% or±1bpm, which is greater
Resolution 1 bpm 
Sensitivity > 200 (uVp-p) 
Differential Input Impedance > 5 MΩ
CMRR Monitor > 105 dB 
Operation > 105 dB 
Diagnosis > 85 dB


Electrode offset potential ±300mV 
Leakage Current < 10 uA
Baseline Recovery < 3 S After Defi. 
ECG Signal Range ±8 m V ( Vp-p )
Bandwidth Surgery 1 ~ 20 Hz
Monitor 0.5 ~ 40 Hz
Diagnostic 0.05 ~ 130 Hz
Calibration Signal 1 (mVp-p), Accuracy:5%
ST Segment Monitoring Range Measure and Alarm -2.0 ~ +2.0 mV

ARR Detecting Type: ASYSTOLE, VFIB/VTAC, COUPLET, BIGEMINY, TRIGEMINY, R ON T, VT>2,BRADY, MISSED BEATS,
PNP, PNC,Alarm Available,Review Available

 

Respiration
Method Impedance between R-F(RA-LL)
Differential Input Impedance > 2.5 MΩ
Measuring Impedance Range 0.3~3 Ω
Base line Impedance Range 200~4 KΩ
Bandwidth 0.1~2.5Hz
Resp. Rate Measuring and Alarm Range
Adult 0 ~ 120 Brpm 
Neo/Ped 0 ~ 150 Brpm 
Resolution 1 Brpm 
Accuracy ±2 Brpm
Apnea Alarm 10 ~ 40 S

NIBP
Method Oscillometric
Mode Manual, Auto, STAT 
Measuring Interval in AUTO Mode 1, 2, 3, 4, 5, 10, 15, 30, 60, 90,120, 180, 240,480 (Min)
Measuring Period in STAT Mode 5 Min
Pulse Rate Range 40 ~ 240 bpm
Alarm Type SYS, DIA, MEAN

Measuring and alarm range Adult Mode
SYS 40 ~ 270 mmHg 
DIA 10 ~ 215 mmHg 

MEAN 20 ~ 235 mmHg
Pediatric Mode
SYS 40 ~ 200 mmHg 
DIA 10 ~ 150 mmHg 
MEAN 20 ~ 165 mmHg
Neonatal Mode
SYS 40 ~ 135 mmHg 
DIA 10 ~ 100 mmHg 
MEAN 20 ~ 110 mmHg

Resolution 1mmHg
Accuracy Maximum Mean error ±5mmHg
Maximum Standard deviation ±8mmHg
Overpressure Protection Adult Mode 297±3 mmHg 
Pediatric Mode 240±3 mmHg 
Neonatal Mode 147±3 mmHg


SpO2
Measuring Range 0 ~ 100 % 
Alarm Range 0 ~ 100 % 
Resolution 1 %
Accuracy 70% ~ 100% 2 %
0% ~ 69% unspecified
Actualization intervalabout 1 Sec. 
Alarm Delay 10 Sec.
Pulse Rate Measuring and Alarm Range 20~300 bpm
Resolution 1bpm
Accuracy ±3bpm

TEMP
Channel 2
Measuring and Alarm Range 0 ~ 50 ℃
Resolution 0.1 ℃
Accuracy ±0.2 ℃
Actualization intervalabout 1 Sec.
Average Time Constant < 10 Sec.

IBP
Label ART, PA, CVP, RAP, LAP, ICP, P1, P2
Measuring and alarm range

ART 0 ~ 300 mmHg 
PA -6 ~ 120 mmHg
CVP/RAP/LAP/ICP-10 ~ 40 mmHg
P1 / P2-10 ~ 300 mmHg
Press Sensor Sensitivity 5 uV/V/mmHg
Impedance 300-3000Ω
Resolution 1 mmHg
Accuracy ±2% or 1mmHg which is greater
Actualization interval about 1 secretary

Standard: ECG, Respiration, NIBP, SpO2, Pulse Rate, Temperature
Optional: Nellcor SpO2, EtCO 2, IBP, Thermal Recorder



t